Temperature control for a Janis Research ST-LN-500_1-4CX LN₂ cryogenic probe station via a Cryo-con 22C controller on COM5.
Start here: FINAL_REPORT.html — the overshoot study,
the settings that fixed it, and the LabVIEW SCPI sequence. Open it in a browser;
it is self-contained and prints cleanly.
LOOP 1:PGAIN 40 proportional gain
LOOP 1:IGAIN 900 integral time, SECONDS (not a gain)
LOOP 1:DGAIN 0 derivative off
LOOP 1:RANGE HI 50 W full scale
LOOP 1:MAXPWR 70 percent
LOOP 1:RATE 1.0 K/min (2.0 works equally well and is twice as fast)
LOOP 1:TYPE RAMPP
Peak overshoot +0.22 K, down from +4.08 K. PID Table 2 in the instrument's
memory now carries these Zone 1 values as well, so RAMPT/TABLE mode is safe
above 300 K.
Send CONTROL before the target setpoint, and send the setpoint last —
writing the setpoint is what arms the ramp. Setting a target while control is OFF
makes the heater slam to its power limit the moment control is switched on.
- Port:
COM5(or assigned USB-serial COM port) - Baud Rate:
57,600(8 Data Bits, 1 Stop Bit, No Parity,\r\nLine Termination) - Hardware Upgrade Note: Upgraded from the factory 9,600 baud to 57,600 baud on the instrument front panel (
System -> Remote -> RS-232 Rate: 57600). - Telemetry Throughput & ADC Limits:
- The Cryo-con 22C has a fixed 24-bit delta-sigma ADC running at 10 Hz (100 ms).
- At 57,600 baud, single SCPI query roundtrips take ~25–30 ms (down from ~120 ms at 9600).
- The Python GUI (
cryocon_gui.py) features an Ultra (0.10 s / 10 Hz) polling option that streams stage temperature, heater output, and setpoints at the full hardware ADC limit with zero lag.
| Path | Contents |
|---|---|
FINAL_REPORT.html / .md |
The report |
OVERSHOOT_TUNING_LOG.md |
Full technical log — every attempt, superseded theories, safety config |
data/ |
26 measured runs + the PID table backup taken before overwriting |
figures/ |
Report figures (regenerate with make_final_figures.py) |
docs/ |
Cryocon manual, project objective, dev notes |
labview/ |
LabVIEW port — build spec, 22C simulator, ActiveX test harness |
archive/ |
Superseded scripts and the earlier report's generators |
data_backup_20260904/ |
Independent backup taken mid-study |
Cryocon-22C_and_Wayne_Kerr/ |
Unified suite: Wayne Kerr 6510B Impedance Analyzer + Cryo-con 22C |
The folder Cryocon-22C_and_Wayne_Kerr/ contains the unified impedance spectroscopy software used for characterizing PMN-0.3PT ($0.70\text{Pb}(\text{Mg}{1/3}\text{Nb}{2/3})\text{O}_3 - 0.30\text{PbTiO}_3$):
-
unified_gui.py: Unified GUI orchestrating simultaneous temperature ramps and multi-frequency dielectric impedance sweeps with live$\varepsilon'(T)$ visualization. -
Latest Data: Subfolder
Cryocon-22C_and_Wayne_Kerr/Data/contains the latest 57-point dataset (300 K to 412 K) for PMN-0.3PT. -
Launcher: Run
Cryocon-22C_and_Wayne_Kerr\run_gui.batorpython Cryocon-22C_and_Wayne_Kerr/unified_gui.py. -
Documentation: See
Cryocon-22C_and_Wayne_Kerr/README.mdandLEARNING_LOG.mdfor full physical and operational details.
| Script | Purpose |
|---|---|
staged_ramp_test.py |
Ramp runner for tests. Sends STOP on every exit path — use this for anything unattended. |
live_view.py |
Live temperature/heater display. Tails the log file, so it does not occupy COM5 while a test runs. |
write_pid_table02.py |
Writes PID Table 2 to instrument NVRAM. Backs nothing up itself — take a copy first. |
make_final_figures.py |
Regenerates every report figure from data/. |
read_full_table.py |
Dumps the PID tables currently in the instrument. |
monitor_only.py |
Read-only monitor. |
cryocon_gui.py |
GUI (run_gui.bat). |
Only one process can hold COM5. Close the GUI and stop any running test before starting another.
- 77–300 K zone is untuned. Its table values are likely wrong for the same units reason as Zone 1, but LN₂ changes the dynamics and there is no data yet. LN₂ is on order — retune once it arrives.
- Over Temperature Disconnect (Channel A, 470 K) is set by hand on the front panel and is not remotely settable or readable. Re-enter it manually if the instrument is ever reset. (Updated from 460 K).